Accurate, simple and cost effective!
      • Controlled Accuracy - Automatic gate allows only 1 tablet or capsule to pass across light beam at a time
      • Small Footprint - lets you place it anywhere you need and easily stored away when not in use (W=10.5" x L=12.5" x H=7.0" Weight=7lbs)
      • Single Counts - Counts into top drawer and transfers to vial. Excess transfers to lower drawer so it can be replaced into stock container.
      • Multiple Counts - Remove top drawer at end of initial count. By pushing "MULTIPLE" button, remaining tablets will stay on top of counter, ready for next count. Ideal for prepackaging.
      • Inventory Counts - By pressing "INV" button, the RX-4 will count up to 1000 and restart at 1 automatically.
      • Available in 110 or 220  voltage requirements
      • Shipping Information: 17” x 13” x 10” - 10lbs
      • NOTE - International Orders May Require Wire Transfer Payment
 
The RX-4 is highly accurate…the automatic gate allows only 1 tablet or capsule to pass by the machine’s electric eye. With its two drawer design, the RX-4 can be used for either single or multiple counts. It can be equipped for either 110 Volt or 220 Volt power requirements. And, the machine’s small footprint lets you place it anywhere you need and easily stored away when not in use.




Accuracy Testing & Results
Testing Date: August 1, 2013 

Single Mode Multi Mode Inventory Mode
(30 pills/cycle) (30 pills/cycle) (100 pills/cycle)
Cycles Tablet 250 250 250
Capsule 250 250 250
Miscounts Tablet 1 * 2 * 0
Capsule 0 0 0
Error Percentage Tablet 0.4% 0.8% 0.0%
Capsule 0.0% 0.0% 0.0%
Accuracy Tablet 99.60% 99.20% 100.00%
Capsule 100.00% 100.00% 100.00%
Overall 99.80% 99.60% 100.00%

* All miscounts occurred as a result of a single extra unit passing through the multi-sensor due to momentum of rotating disc. Each excess tablet was counted and indicated on the digital display therefore it is not regarded as a true miscount because it was detected by the multi-sensor. For example, on a test count of 30, an extra tablet passed the multi-sensor as the rotating disc came to a stop. The extra tablet was counted and shown on the digital display as a count of 31.

Testing was performed on the RX-4, 110 volt. Tablets and capsules were tested separately. Testing was performed in each mode as follows:
  • Single: Performed 250 cycles of 30 tablets and 250 cycles of 30 capsules for a total of 500 cycles.
  • Multi: Performed 250 cycles of 30 tablets and 250 cycles of 30 capsules for a total of 500 cycles.
  • Inventory: Performed 250 cycles of 100 tablets and 250 cycles of 100 capsules for a total of 500 cycles.

Results: The RX-4 performed at an overall accuracy level of 99.8% with no true miscounts occurring. Additionally, no miscounts occurred during testing with capsules. No "undercounts" occurred.

Notes: The counting speed of the unit is directly attributable to the size of the pill being counted…smaller = faster. A typical, round, aspirin-size pill goes through at about 500/minute. Pill size can range anywhere from about 3mm dia. to large 000 size capsules.  All new machines are thoroughly tested prior to shipping to ensure 100% accuracy and are backed by a full one-year warranty on parts and labor.